6. Warning and Error Messages
6.1 Warnings
Warnings indicate a condition that requires user attention but does not necessarily stop operation.
Table 6-1 BactoSwitch warnings
| ERROR CODE | NAME | DESCRIPTION | CAUSES / WHAT TO DO |
|---|---|---|---|
| W15 | BactoSwitchLowFlow | The input flow is lower than the advised value. Measurement precision could be degraded. | Typical causes: inlet pressure below specification, partially closed upstream valve, or installation creating back-pressure at the drain. Recommended action: • In Test BactoSwitch, open the affected line alone and adjust it to 150 – 400 mL/min using the flow rate adjustment knob (see Line testing and flow setting). • Confirm that the inlet pressure is > 0.1 bar and Δh > 0.5 m (see Figure 4-2). • Correct any upstream restrictions. |
6.2 Errors
Table 6-2 BactoSwitch errors
| ERROR CODE | NAME | DESCRIPTION | CAUSES / WHAT TO DO |
|---|---|---|---|
| E60 | BactoSwitchFlowTooLowError | The measured flow on the selected input is below the acceptable range for single-line operation. | Typical causes: low inlet pressure; upstream restriction; drain installation causing back-pressure; valve manually closed. Recommended action: • In Test BactoSwitch, open the affected line alone and adjust it to 150 – 400 mL/min using the flow rate adjustment knob (see Line testing and flow setting). • Confirm that the inlet pressure is > 0.1 bar and Δh > 0.5 m (see Figure 4-2). • Correct any upstream restrictions. • Check that the On/Off valve is open. |
| E61 | BactoSwitchDeviceNotFoundError | The BactoSense does not detect the BactoSwitch. | Typical causes: power not applied to the BactoSwitch; USB-C cable not connected or not screwed on both devices. Recommended action: Connect the power supply to the BactoSwitch and secure the USB-C cable with the screw collars on both the BactoSwitch and the BactoSense. The device is detected automatically when connected. |
| E62 | BactoSwitchTimeoutError | A communication attempt with the BactoSwitch exceeded the expected time. | Typical causes: temporary communication issue on the USB-C link; loose or unscrewed connectors. Recommended action: Check that the USB-C cable is seated and screwed on both ends and that the BactoSwitch is powered, then retry the operation. |
| E63 | BactoSwitchCommunicationError | A communication error occurred between the BactoSense and the BactoSwitch. | Typical causes: unstable or incorrect USB-C connection; cable not secured. Recommended action: Inspect and re-secure the USB-C cable on both devices (use the screw collars), verify power to the BactoSwitch, and reboot both devices. |
If issues persist after applying the steps above, contact your local service partner (see Contact).
For the warnings and errors of the BactoSense itself, see Warnings, errors and troubleshooting in the BactoSense user manual.