16. Warnings, Errors and Troubleshooting

If the instrument encounters a problem, it enters a different state, depending on the severity of the issue. There are warnings, noncritical errors and critical errors, each having different implications and requiring different actions.

16.1 Warnings​

Warnings appear when unusual behaviour is detected during an otherwise successful measurement. They can indicate reduced accuracy of the measurement results or indicate impending errors. Unlike errors, warnings do not prevent the instrument from functioning, but users should pay attention to them as they can indicate sources of inaccuracies.

WARNINGIMAGE
When unusual behavior is detected, a warning is shown next to the measurement results. More information on the warning can be found by clicking on the corresponding warning entry next to the measurement results or directly in the warning log.BactoSense — High SSC baseline warning

The following warning messages can be displayed:

Table 16-1 Warnings

ERROR CODENAMEDESCRIPTIONCAUSES / WHAT TO DO
W01Cartridge expiredCartridge reagents have expired. Cell counts may be underestimated. Please replace the cartridge as soon as possible.• Please replace the cartridge as soon as possible.
W02-04High baseline signal in FL1 / FL2 / SSCBaseline signals are too high. This is usually due to bubbles or particles in suspension in the sample. Measurement accuracy could be affected.• Check the inlet tubing for any signs of deposits.
• Check the water connections, be sure that they are all tight (no air entry).
• If the warning persists, please perform cleaning of instrument with the cleaning kit.
W05Incubator temperature off targetTemperature during incubation was more than 2°C off target. Cell counts may be underestimated. The incubator may be defective. If the problem persists, please call customer service.• If the problem persists, please contact customer service.
W06No events detectedNo events detected during the measurement. The photodetectors are probably defective. If the problem persists, please call customer service.• If the problem persists, please contact customer service.
W07Cell count out of rangeCell count is above 2'000'000 cells/mL, the specification limit of the device. Please dilute the sample to ensure measurement accuracy.• Please dilute the sample to ensure accuracy of the measurement.
W08Air bubbles detectedThe instrument detected more than 10% volume in bubbles. Measurement precision could suffer.• Check sampling device is installed properly.
• Check the sample for bubbles.
• If using online sampling device, check water is supplied at correct pressure.
W10Enclosure too coldEnclosure is too cold, rapid heating has been activated before continuing protocol.• Instrument cannot safely operate if internal temperature is too cold.
• Rapid heating attempts to heat the enclosure to ensure safe operations.
• Measurements will start once internal temperature is sufficient.
W11Service requiredService of the instrument is required. Measurement precision can no longer be guaranteed. Continued operation can lead to leaks, which can damage the instrument.• Next service date is overdue.
• Valve and plunger have reached end of life.
• Please contact your service representative.
W13Laser end of lifeLaser has reached end of life.• Please contact your service representative.
W14Non-critical power cutThe instrument was powered off unexpectedly without a cleaning cycle. Measurement precision could be degraded.• Ensure the cleanliness of the instrument.
• If the issue occurs frequently, install an uninterruptible power supply

16.2 Non-critical errors​

Non-critical errors prevent a measurement from terminating successfully, but do not prevent the instrument from running another measurement afterwards. These errors do not require human intervention. Some non-critical errors are promoted to critical errors if they repeat three times.

NON-CRITICAL ERRORS
The protocol stops. The cause of the error is usually fixed by repeating the analysis or waiting.
• Another measurement can immediately be started. If it completes successfully, the error state is cleared.
• More information on the error can be found in the error log, by clicking on the corresponding error entry.
BactoSense — Non-critical empty sample error

The following non-critical error messages can be displayed:

Table 16-2 List of non-critical errors.

ERROR CODENAMEDESCRIPTIONCAUSES / WHAT TO DO
E01Cartridge door openCartridge door is open and prevents measurements from running.• Close door and retry the measurement.
E08Enclosure too dampEnclosure humidity is too high for safe operation.• Replace desiccant bag, by unscrewing the large cap on the left-hand-side of the instrument (see Replacement of the desiccant bag).
• If the problem persists, please contact customer service.
E09System overheatedInside temperature is too high for safe operation.• Reduce ambient temperature or increase measurement interval to allow instrument to cool down.
E14System overheatedLaser temperature is too high for safe operation.• Reduce ambient temperature or increase measurement interval to allow instrument to cool down.
E15Empty sample errorUnderpressure detected during mix or dispense.• Missing sample.
• Air in the input.
• Leak in the system.
• Please execute the Clean Optics protocol (see Overview of protocols in manual mode).
E18Under-pressure detectedUnderpressure detected during dispense.• Missing sample.
• Air in the input.
• Leak in the system.
• Please execute the Clean Optics protocol (see Overview of protocols in manual mode).
E19System overheatedThe incubator chamber temperature is too high for safe operation.• Reduce ambient temperature, or increase measurement interval to allow instrument to cool down.
E33Processing memory overflowSignal processing circuits memory has overflowed.• Please execute the Clean Optics protocol (see Overview of protocols in manual mode).
• Retry the measurement.
E42Enclosure too cold despite heatingEnclosure temperature remains too cold after heating procedure.• Ambient temperature is less than 5 °C (41 °F).
• Increase ambient temperature.
E44Pump module startup errorThe pump module failed to power on correctly.• Simply retry the measurement.
E45Low valid volume ratio errorThe valid volume ratio is too low. This can be caused by an excessively high bacterial concentration in the water or by turbidity/debris.• The accuracy of the measurement is probably very low. Repeat the measurement while diluting the sample with clear water.
E46Cartridge is still busyThe cartridge was still initializing when the measurement started.• Non-critical error. Repeat the measurement. If the problem persists, contact a service technician.
E48Microfluidic subsystem busyThe microfluidic controller was performing another task when the measurement was taken.• Non-critical error. Repeat the measurement. If the problem persists, contact a service technician.
E49GPIO Timeout errorThe processing board is waiting for another module to respond.• Non-critical error. Repeat the measurement. If the problem persists, contact a service technician.
E50Pump motion blockedThe motion of the pump is blocked.• Non-critical error. Repeat the measurement. If the problem persists, contact a service technician.
E51Unable to get temperature reading. The control loop is thus disabled.Unable to get temperature reading. The control loop is thus disabled.• Non-critical error. Repeat the measurement. If the problem persists, contact a service technician.
E52Cartridge valve motion blockedThe motion of the cartridge valve is blocked.• Non-critical error. Repeat the measurement. If the problem persists, contact customer service.
E55The acquisition processor cannot be initialized.The acquisition processor cannot be initialized. This can be caused by an acquisition error or a firmware corruption.• Reboot the instrument.
• If the problem persists, contact customer service.
E56No events detected during acquisitionNo events were detected during the acquisition. This can be caused by a high turbidity of the sample or defective detectors.• Clean the sampling device using the Washstation, then perform a measurement of clean water.
• If the problem persists, contact customer service.
E57Cartridge busyCommunication with the cartridge failed.• Non-critical error. Repeat the measurement.
• If the problem persists, contact customer service.
E58Cartridge busyCommunication with the cartridge failed.• Check the electronic connector between the cartridge and the instrument.
• Repeat the measurement.
• If the problem persists, contact customer service.
E59Unable to read valve positionUnable to read valve position. The sensor cable may be disconnected.• Please contact customer service.

16.3 Critical errors​

Critical errors interrupt the workflows of the instrument and require human intervention. The error source needs to be eliminated and the instrument set to running conditions again. Critical errors need to be cleared.

CRITICAL ERRORSIMAGE
If a critical error occurs during operation, it has the following effects:
• The protocol immediately stops.
• The instrument goes into a critical error state, and manual intervention is needed before any new protocol can be launched.
• The cause of the error must be solved by an operator. The errors can then be manually cleared from the Error log, as described in Error log and Clearing errors.
BactoSense — Critical missing cartridge error

The following critical error messages can be displayed:

Table 16-3 List of critical errors.

ERROR CODENAMEDESCRIPTIONCAUSES / WHAT TO DO
E00General errorUnhandled error.• Please contact customer service
E02Empty cartridgeCartridge is empty.• Replace cartridge (see Change cartridges)
E04Missing cartridge or door openCartridge is missing.• Insert a cartridge (see Installing a cartridge)
E05Full waste bagCartridge waste bag is full.• Replace cartridge (see Change cartridges)
E06Cartridge communication impossibleCartridge is disconnected, door is open, or cartridge electronics are damaged.• Make sure the cartridge’s electronic cable is connected properly, and that the door is properly closed.
• If that doesn’t help, replace the cartridge and inform customer service.
E07Storage disk fullInsufficient storage space to continue operation.• Delete old measurements and try again (see Clearing of disk space)
• This will delete the measurements permanently from the device.
• Export measurements as backup solution.
E12FPGA initialization errorFailed to initialize the signal processing chip.• Reboot the instrument.
E16Processing errorAn error was discovered while processing the signals.• The accuracy of the latest result is not guaranteed. Please repeat the analysis.
E17Over-pressure detectedOverpressure detected during dispense.• Likely cause: The output filter is probably clogged.
• Optical flow cell blocked.
• Please call customer service.
E20Abrupt shutdownProtocol interrupted for unknown reasons, usually due to a power cut.• Usually indicates a power cut, or other external interruption to the measurement.
• Perform a Clean Optics and Clean Sampling Device protocol before starting a new measurement (see Overview of protocols in manual mode).
E23Temperature sensor communication errorCommunication with temperature sensor on IO board failed.• Please contact customer service.
E24Laser communication errorCommunication with optical unit failed.• Please contact customer service.
E25Pump communication errorCommunication with syringe-pump module impossible.• Please contact customer service.
E26Power supply communication errorCommunication with power supply module impossible.• Please contact customer service.
E27Pump communication timeoutCommunication with pump module timed out.• The cartridge door is open.
• A cable is disconnected.
• A circuit board is damaged, either in the pump or cartridge.
E28Empty dye supplyDye tube connected to cartridge is empty.• The tube is disconnected.
• Cartridge is defective.
• Repeat the cartridge change procedure, making sure that all tubes are properly connected.
E29Empty rinse fluid supplyRinse fluid tube connected to cartridge is empty.• The tube is disconnected.
• Cartridge is defective.
• Repeat the cartridge change procedure, making sure that all tubes are properly connected.
E30Empty bleach supplyBleach tube connected to cartridge is empty.• The tube is disconnected.
• Cartridge is defective.
• Repeat the cartridge change procedure, making sure that all tubes are properly connected.
E31Cartridge electronics failureCartridge level can no longer be determined accurately.• Replace cartridge and contact customer service.
E32Incompatible cartridgeCartridge type is incompatible with this instrument or software.• Make sure that you have the correct cartridge type for this instrument.
E33Processing memory overflowSignal processing circuits memory has overflowed.• Please execute the Clean Optics protocol (see Overview of protocols in manual mode).
• Retry the measurement.
E38Cartridge not ready errorThe new cartridge has not been initialized correctly.• Execute the “Cartridge Change” procedure (see Change cartridges).
E39Laser current limit reachedThe laser has reached its upper current limit; indicating end of life.• The laser needs to be replaced. Contact a service technician.
E40Gate set not foundReference to the set of gates selected for the measurement cannot be found in the instrument settings.• The set of gates has been deleted.
• Change the set of gates and retry the measurement.
E41Gate set incompatibleThe set of gates selected for the measurement cannot be used with this cartridge.• Correct cartridge change procedure was not followed.
• Incorrect digital input configuration.
• Change the set of gates and retry the measurement.
E43External temperature too lowAmbient temperature is outside of the instrument specifications.• Increase ambient temperature.
• Trying to operate or store the instrument below 5 °C (41 °F) can cause damage to the device.
E47Unable to find main reference magnet during homingDuring initialization, the valve did not home properly.• Please contact customer service.
E53Unable to communicate with motorThe communication to the motor controller of the mixer is defective.• Please contact customer service.
E54Unable to get the laser temperatureThe NTC sensor in the optical module must be defective or disconnected.• Please contact customer service.

16.4 Logs​

The instrument records errors, warnings, alarms and user activity in logs. On the Home screen, press Settings, then Logs. The Logs menu contains the following entries:

ENTRYCONTENTAVAILABLE TO
Error logCritical and non-critical errors, and the events that cleared them. See Error log.Basic, Advanced and Admin users
Warning logWarnings raised during measurements. See Warning log.Basic, Advanced and Admin users
Alarm logAlarms triggered by measurement results, including BactoCloud alarms. See Alarm log.Basic, Advanced and Admin users
Access logFailed login attempts. See Access log and audit trail.Admin users, with the Management Package
Audit trailOperations and setting changes by users and the instrument. See Access log and audit trail.Admin users, with the Management Package
Results (manual mode), Results (Online mode), Validation resultsMeasurement results of each mode. See Filtering measurements and Results and Data Analysis.Basic, Advanced and Admin users

The newest entries are listed first. The Error log, Warning log, Alarm log and Audit trail show the Displayed time span: in the action bar. By default, it runs from the oldest to the newest entry; the Audit trail shows the last seven days. Press the time span to show another period (see Changing the displayed time span).

16.4.1 Error log​

BactoSense — Error log

Figure 16-1 : Error log

①Date and time of the event②Type of the event. Only the newest entry shows its symbol in color; older entries show it in grey.
BactoSense — Critical error log entryCritical error
BactoSense — Non-critical error log entryNon-critical error
BactoSense — Log entry: errors clearedErrors cleared
③Code and short description of the error or event. The number in brackets shows how often the error has occurred since errors were last cleared.④Clear errors button (available to Basic, Advanced and Admin users when the instrument is idle). See Clearing errors.
⑤Displayed time span. Press it to change the period.

The row color shows the type of each entry: red for critical errors, yellow for non-critical errors and blue for events that cleared errors. Error codes start with E and are listed in Non-critical errors and Critical errors. Two events record that errors were cleared:

  • C01 – Errors cleared automatically: after a non-critical error, the instrument completed a protocol successfully and considers the earlier errors resolved.
  • C02 – Errors cleared manually: a user pressed Clear errors (see Clearing errors).

If no error was recorded in the displayed time span, No errors in this date range is shown.

View event details

Select an entry to open a page with more information on the cause of the error and recommended resolution steps. Press Close to return to the log.

16.4.2 Warning log​

The Warning log lists the warnings raised during measurements. Each row shows:

  • Date: date and time of the warning.
  • Description: warning code and name. The codes are listed in Warnings.
  • Context: where the warning occurred: Manual (manual mode), Online (Online mode), Maintenance or Validation.

Select a warning to view its description. Press Close to return to the log. If no warning was recorded in the displayed time span, No warnings in this date range is shown.

16.4.3 Alarm log​

The Alarm log lists the alarms triggered by measurement results. It includes the instrument's own alarms (see Alarm configuration) and, when the instrument is connected to BactoCloud, the alarms raised by BactoCloud alarm rules (see BactoCloud). Each row shows:

  • Date: date and time of the measurement that triggered the alarm.
  • Name: a bell symbol and the alarm name. The bell is red for instrument alarms. For BactoCloud alarms, its color shows the severity: red for Critical, orange for Warning and blue for Info.
  • Alarm: the alarm source, condition and threshold, for example Cartridge Remaining < 10 %.

BactoCloud alarms also show a cloud symbol at the end of the row. To acknowledge a BactoCloud alarm after reviewing it, press the check button beside the cloud symbol; acknowledged alarms show a check mark instead. Acknowledge all acknowledges every unacknowledged BactoCloud alarm in the list and is only available while at least one remains. Acknowledgement is recorded in BactoCloud. It documents the review but does not resolve the cause of the alarm.

Select an alarm to open the measurement result that triggered it.

16.4.4 Access log and audit trail​

The Access log and the Audit trail are part of the Management Package (see Management package) and are available to Admin users. Without the package, pressing either button shows information about the Management Package instead.

Access log opens the Failed login attempts list for the last 30 days. Each row shows the Date, the User ID, the user's Role and the number of failed login attempts by that user on that day. If there were none, No failed login attempts is shown.

Audit trail lists relevant operations and setting changes made by users or by the instrument. Each entry shows the Date, the User and the Event and, where they apply, the changed Parameter with its Old value and New value. The audit trail is also included when exporting all data to a USB drive and can be viewed in the web user interface.

16.4.5 Changing the displayed time span​

The following procedure applies to the Error log, Warning log, Alarm log and Audit trail.

WORK STEPADDITIONAL INFO / IMAGE
1.Open the log and press the time span button next to Displayed time span:.The Define time span screen opens.
2.Tap the field under Start date and select the first day in the calendar. Tap the time field below it and set the hour and minute. Set End date and its time in the same way.The end cannot be earlier than the start.
3.Press OK.The log lists only the entries within the selected period.

To restore the default time span, press Reset on the Define time span screen, then press OK.

16.5 Clearing errors​

Critical errors must be cleared manually. Make sure to first solve the origin of the problem, then manually clear the error with the steps below. Please note that the procedure below does not automatically solve the root cause of the error.

WORK STEPADDITIONAL INFO / IMAGE
1.On the Home screen, press Settings to open the settings menu.
2.Press Logs, then Error log.
3.Make sure you have resolved the problem leading to the displayed error(s). Then press Clear errors.The button is available to Basic, Advanced and Admin users while no protocol is running.
BactoSense — Clear errors control
4.A C02 – Errors cleared manually event is added at the top of the log, and the symbols of the earlier errors turn grey. The instrument is ready to analyze again.

16.6 Automatic self-check​

The automatic self-check tool checks that each component of the BactoSense is functioning correctly and identifies those that are not. It can be used to identify problems with the instrument.

WORK STEPADDITIONAL INFO / IMAGE
1.As an Advanced or Admin user, press Settings on the Home screen to open the settings menu.
2.Press the Maintenance button.
3.Press the Self-check button.
4.
Additional information

Press the Rerun button. Once execution has terminated, the instrument shows a list of all tests, with a green PASS or red FAIL indicator. Some tests may be skipped: if the system cannot communicate with the cartridge, all the tests that require cartridge communication are skipped.

BactoSense — Self-check results

16.7 Operation at low temperature​

The instrument must be operated within a specific temperature range to prevent damage and ensure reliable and reproducible measurement results (see Technical data). At very low and very high ambient temperatures (below 5 °C (41 °F) or above 30 °C (86 °F)), protocols are not executed. At low ambient temperatures, the instrument automatically warms up prior to a measurement to reach the required temperature. During this heating process, the instrument status shows “Heating...”. If the target temperature cannot be reached after several hours, the protocol exits with an error (table 16-2, E42).

Allow time for heating

Measurements can be delayed if extensive heating is necessary. To avoid this, keep the instrument temperature above 10 °C (50 °F).

16.8 Troubleshooting​

The following table provides an overview of issues and resolutions. If the issue cannot be resolved, please contact your local service representative or bNovate customer service. Contact information for bNovate customer service can be found in Contacting customer service.

Table 16-4 Troubleshooting instructions

OBSERVATIONINSTRUCTION
Nothing on display• Check whether the BactoSense is connected to power and switched on.
Unresponsive or flickering user interface• Potential electromagnetic interference from nearby machine (e.g. pump). Relocate BactoSense.
Error message on the display• Respond to the error message as described in this chapter.
The measurements appear to be wrong• Ensure that the sample and operating conditions described in Technical data are met.
• Check whether the instrument is correctly installed according to Installation.
• Ensure that the servicing duties have been performed according to the servicing schedules described in Periodic preventive maintenance by service professional and Maintenance by the user.
• Perform an instrument cleaning using the Cleaning Kit.
• Perform an instrument validation using the Validation Kit.