3. Instrument Overview
3.1 BactoSense overview
Figure 3-1 BactoSense overview
| 1 | Handle | 2 | Wall mount |
| 3 | Desiccant bag compartment cap | 4 | Online sampling device |
| 5 | Manual sampling device | 6 | Touchscreen graphical user interface |
| 7 | Cartridge compartment door | 8 | USB sockets |
| 9 | Power button | 10 | Input connector |
| 11 | Output connector | 12 | Ethernet/Modbus TCP interface |
| 13 | Power connector | 14 | USB-C socket |
An overview of the cartridge can be found in Cartridge overview.
3.1.1 Type plate of the BactoSense
The type plate of the BactoSense can be found on the right-hand side of the instrument.
Figure 3-2 Type plate of the BactoSense.
| 1 | Manufacturer | 2 | Product name |
| 3 | Serial number | 4 | Barcode |
| 5 | Manufacturing date | 6 | Country of origin |
| 7 | Voltage and power requirements (see Connecting the BactoSense to power) | 8 | CE mark |
| 9 | UKCA mark | 10 | Compliance with Canadian EMC and interference standards |
| 11 | Read the instructions and operate with caution | 12 | Disposal information |
3.1.2 Type plate of the BactoSense power supply
The type plate of the BactoSense power supply can be found on the power supply and has the following elements:
Figure 3-3 Type plate of the BactoSense power supply.
- Manufacturer
- Product name
- Part number
- Manufacturing date
- Service voltage and frequency
- Power consumption
- CE mark
- UKCA mark
- Compliance with Canadian EMC and interference standards
- Read the instructions and operate with caution.
- Disposal information
3.2 Technical data
Table 3-1 Technical data and specifications
| Property | Specification |
|---|---|
| GENERAL | |
| Operating environment | Indoor and protected outdoor use |
| Operating altitude | Max. 2000 m (6600 ft) above sea level |
| Operating temperature | 5 °C – 30 °C (41 °F – 86 °F) |
| Operating humidity | 10 % – 90 % RH |
| Storage temperature | 15 °C – 25 °C (59 °F – 77 °F) |
| Storage humidity | 10 % – 70 % RH |
| Ventilation requirements | None |
| Sound pressure level | < 64 dBA |
| Protection classes | BactoSense: IP65 Power supply: IP67 |
| PHYSICAL | |
| Dimension (W x D x H) | 361 x 270 x 424 mm (14.2 x 10.6 x 16.7“) |
| Weight | With cartridge 14.3 kg (31.5 lbs) Without cartridge 11.1 kg (24.5 lbs) |
| Power supply dimension (W x D x H) | 230 x 80 x 65 mm (9.0 x 3.1 x 2.6“) |
| Power supply weight | 0.9 kg (2.0 lbs) |
| ELECTRICAL | |
| Installation category | II |
| Pollution degree | 2 |
| Power supply | Power supply unit: 100 – 240 VAC +/-10 %, 50 / 60 Hz, 1.4 A BactoSense: 18 – 24 VDC, 60 W (see Connecting the BactoSense to power) |
| Power consumption | Avg. 20 W, max. 60 W |
| Recovery after power interruption | Configurable auto-restart |
| INTERFACES, CONNECTIVITY AND REMOTE ACCESS | |
| USB | 2 x USB 3.1, 5 V 1 A per port 1 x USB-C |
| Ethernet | 1 x RJ45 Gigabit Ethernet |
| Ethernet protocols and options | Web interface Remote desktop (VNC) Modbus TCP HTTP / HTTPS FTP |
| Digital inputs | 4 x digital inputs, optocoupler input: forward voltage 1.2 VDC with 3 kΩ in series, cathode connected to COMMON, max. 30 mA, max. 50 mW |
| Analogue and digital outputs | Analogue outputs: 2 x 0/4 – 20 mA, recommended 250 Ω load Digital outputs: 4 x digital outputs, freely configurable, SSR output: max. 48 VDC, max. 300 mW, ON-resistance 2 Ω |
| Display | WVGA, 7.0″ capacitive touchscreen |
| Mobile connectivity | BactoLink connectivity module (optional) |
| Advanced data analysis | FCS 3.1 file export |
| Data storage capacity | 128 GB |
| MEASUREMENT | |
| Measuring principle | Flow cytometry |
| Detection range | 0 – 5´000´000 cells/mL |
| Quantification range | 1000 – 2‘000‘000 cells/mL for TCC 100 – 2‘000‘000 cells/mL for ICC, ACC, CBC, NBC, EBC |
| Laser type and wavelength | Laser diode 488 nm |
| Fluorescence | 530/50 nm FL1 715 LP nm FL2 |
| Forward scatter (FSC) | 488 nm, instrument dependent |
| Side scatter (SSC) | 488 nm, instrument dependent |
| Result parameters | Primary result parameters: TCC, ICC, ACC, CBC, NBC, EBC Secondary result parameters, cartridge and configuration-dependent: HNAP, HNAC, LNAC, HACC, Gate+ |
| Automatic measuring interval | Min.: 30 min for TCC, ICC, 45 min for ACC, 60 min for CBC, NBC, EBC Max.: 6 hours |
| SAMPLING | |
| Sampling type | Online or manual |
| Online sampling device | Flow through device: 200 – 400 mL/min (0.05 – 0.10 gal/min) Pressure in device: Max. 0.5 bar (7.3 psi) Aspirated volume: 260 μL Measured volume: 90 μL Inlet tubing • Material: PFA (transparent) • Outer diameter: 6.35 mm (1/4”) • Inner diameter: 4.35 mm • Max. length: 2 m (6.5 ft) Outlet tubing • Material: FEP (blue) • Outer diameter: 6.35 mm (1/4”) • Inner diameter: 4.83 mm • Max. length: 2 m (6.5 ft) |
| Manual sampling device | Sample container: 5 mL screw-cap tube Sample volume: 2 – 5 mL Aspirated volume: 260 μL Measured volume: 90 μL |
| Sample conditions | Temperature: 5 °C – 40 °C (41 °F – 104 °F) Turbidity: 0 – 10 NTU, FNU Conductivity: 0 – 100'000 µS/cm at 20 °C for TCC, ICC, ACC 50 – 100'000 µS/cm at 20 °C for CBC, NBC, EBC pH-value: 5 – 12 for TCC, ICC, ACC 6 – 12 for CBC, NBC, EBC Chlorine: max. 3 mg/L Iron: max. 200 μg/L Manganese: max. 50 μg/L |
| CARTRIDGE | |
| Types | TCC, Total Cell Count ICC, Intact Cell Count ACC, Active Cell Count CBC, Coliform Bacteria Count NBC, Enterococcus Bacteria Count EBC, E. coli Bacteria Count Compatibility BactoSense model dependent |
| Capacity | Up to 1000 measurements for TCC, ICC Up to 700 measurements for ACC, CBC, NBC, EBC |
| Validity | 9 months for TCC, ICC 6 months for ACC, CBC, NBC, EBC |
| Storage temperature | 15 °C – 25 °C (59 °F – 77 °F) ideally at 15 °C (59 °F) |
| Storage humidity | 20 % – 60 % RH |
| MAINTENANCE | |
| Preventive maintenance | After 3 cartridges, at least yearly. |
| Calibration | Factory calibration only |
| COMPLIANCE | |
| Applied standards and certifications | Please refer to Compliance and applied standards. |