3. Instrument Overview

3.1 BactoSense overview​

Original manual illustration, printed page 7

Figure 3-1 BactoSense overview

1Handle2Wall mount
3Desiccant bag compartment cap4Online sampling device
5Manual sampling device6Touchscreen graphical user interface
7Cartridge compartment door8USB sockets
9Power button10Input connector
11Output connector12Ethernet/Modbus TCP interface
13Power connector14USB-C socket

An overview of the cartridge can be found in Cartridge overview.

3.1.1 Type plate of the BactoSense​

The type plate of the BactoSense can be found on the right-hand side of the instrument.

Original manual illustration, printed page 8

Figure 3-2 Type plate of the BactoSense.

1Manufacturer2Product name
3Serial number4Barcode
5Manufacturing date6Country of origin
7Voltage and power requirements (see Connecting the BactoSense to power)8CE mark
9UKCA mark10Compliance with Canadian EMC and interference standards
11Read the instructions and operate with caution12Disposal information

3.1.2 Type plate of the BactoSense power supply​

The type plate of the BactoSense power supply can be found on the power supply and has the following elements:

Original manual illustration, printed page 9

Figure 3-3 Type plate of the BactoSense power supply.

  1. Manufacturer
  2. Product name
  3. Part number
  4. Manufacturing date
  5. Service voltage and frequency
  6. Power consumption
  7. CE mark
  8. UKCA mark
  9. Compliance with Canadian EMC and interference standards
  10. Read the instructions and operate with caution.
  11. Disposal information

3.2 Technical data​

Table 3-1 Technical data and specifications

PropertySpecification
GENERAL
Operating environmentIndoor and protected outdoor use
Operating altitudeMax. 2000 m (6600 ft) above sea level
Operating temperature5 °C – 30 °C (41 °F – 86 °F)
Operating humidity10 % – 90 % RH
Storage temperature15 °C – 25 °C (59 °F – 77 °F)
Storage humidity10 % – 70 % RH
Ventilation requirementsNone
Sound pressure level< 64 dBA
Protection classesBactoSense: IP65
Power supply: IP67
PHYSICAL
Dimension (W x D x H)361 x 270 x 424 mm (14.2 x 10.6 x 16.7“)
WeightWith cartridge 14.3 kg (31.5 lbs)
Without cartridge 11.1 kg (24.5 lbs)
Power supply dimension (W x D x H)230 x 80 x 65 mm (9.0 x 3.1 x 2.6“)
Power supply weight0.9 kg (2.0 lbs)
ELECTRICAL
Installation categoryII
Pollution degree2
Power supplyPower supply unit:
100 – 240 VAC +/-10 %, 50 / 60 Hz, 1.4 A
BactoSense:
18 – 24 VDC, 60 W (see Connecting the BactoSense to power)
Power consumptionAvg. 20 W, max. 60 W
Recovery after power interruptionConfigurable auto-restart
INTERFACES, CONNECTIVITY AND REMOTE ACCESS
USB2 x USB 3.1, 5 V 1 A per port
1 x USB-C
Ethernet1 x RJ45 Gigabit Ethernet
Ethernet protocols and optionsWeb interface
Remote desktop (VNC)
Modbus TCP
HTTP / HTTPS
FTP
Digital inputs4 x digital inputs, optocoupler input: forward voltage 1.2 VDC with 3 kΩ in series, cathode connected to COMMON, max. 30 mA, max. 50 mW
Analogue and digital outputsAnalogue outputs:
2 x 0/4 – 20 mA, recommended 250 Ω load
Digital outputs:
4 x digital outputs, freely configurable,
SSR output: max. 48 VDC, max. 300 mW,
ON-resistance 2 Ω
DisplayWVGA, 7.0″ capacitive touchscreen
Mobile connectivityBactoLink connectivity module (optional)
Advanced data analysisFCS 3.1 file export
Data storage capacity128 GB
MEASUREMENT
Measuring principleFlow cytometry
Detection range0 – 5´000´000 cells/mL
Quantification range1000 – 2‘000‘000 cells/mL for TCC
100 – 2‘000‘000 cells/mL for ICC, ACC, CBC, NBC, EBC
Laser type and wavelengthLaser diode 488 nm
Fluorescence530/50 nm FL1
715 LP nm FL2
Forward scatter (FSC)488 nm, instrument dependent
Side scatter (SSC)488 nm, instrument dependent
Result parametersPrimary result parameters:
TCC, ICC, ACC, CBC, NBC, EBC
Secondary result parameters, cartridge and configuration-dependent:
HNAP, HNAC, LNAC, HACC, Gate+
Automatic measuring intervalMin.: 30 min for TCC, ICC, 45 min for ACC, 60 min for CBC, NBC, EBC
Max.: 6 hours
SAMPLING
Sampling typeOnline or manual
Online sampling deviceFlow through device:
200 – 400 mL/min (0.05 – 0.10 gal/min)
Pressure in device: Max. 0.5 bar (7.3 psi)
Aspirated volume: 260 μL
Measured volume: 90 μL
Inlet tubing
• Material: PFA (transparent)
• Outer diameter: 6.35 mm (1/4”)
• Inner diameter: 4.35 mm
• Max. length: 2 m (6.5 ft)
Outlet tubing
• Material: FEP (blue)
• Outer diameter: 6.35 mm (1/4”)
• Inner diameter: 4.83 mm
• Max. length: 2 m (6.5 ft)
Manual sampling deviceSample container: 5 mL screw-cap tube
Sample volume: 2 – 5 mL
Aspirated volume: 260 μL
Measured volume: 90 μL
Sample conditionsTemperature: 5 °C – 40 °C (41 °F – 104 °F)
Turbidity: 0 – 10 NTU, FNU
Conductivity: 0 – 100'000 µS/cm at 20 °C for TCC, ICC, ACC
50 – 100'000 µS/cm at 20 °C for CBC, NBC, EBC
pH-value: 5 – 12 for TCC, ICC, ACC
6 – 12 for CBC, NBC, EBC
Chlorine: max. 3 mg/L
Iron: max. 200 μg/L
Manganese: max. 50 μg/L
CARTRIDGE
TypesTCC, Total Cell Count
ICC, Intact Cell Count
ACC, Active Cell Count
CBC, Coliform Bacteria Count
NBC, Enterococcus Bacteria Count
EBC, E. coli Bacteria Count
Compatibility BactoSense model dependent
CapacityUp to 1000 measurements for TCC, ICC
Up to 700 measurements for ACC, CBC, NBC, EBC
Validity9 months for TCC, ICC
6 months for ACC, CBC, NBC, EBC
Storage temperature15 °C – 25 °C (59 °F – 77 °F) ideally at 15 °C (59 °F)
Storage humidity20 % – 60 % RH
MAINTENANCE
Preventive maintenanceAfter 3 cartridges, at least yearly.
CalibrationFactory calibration only
COMPLIANCE
Applied standards and certificationsPlease refer to Compliance and applied standards.