System Suitability Test Kit for BactoSense PURE

General information​

According to United States Pharmacopeia USP <1058>, the System Suitability Test (SST) verifies that a system performs in accordance with the criteria set forth in the test procedure. The SST Kit for BactoSense PURE is a reagent kit used to conduct this performance verification.

Prerequisites:

  • BactoSense PURE with the manual sampling device installed.
  • SST Kit stored under adequate conditions and not expired.
  • A personal instrument login with Advanced or Admin permissions.

Store the kit refrigerated at 2–8 °C (35–47 °F), away from light. Under these conditions, the shelf-life is one year. The kit provides 10 tests.

Safety information​

Wear protective gloves

In case of skin contact, rinse immediately and abundantly with fresh water. In case of pain or irritation, consult a doctor.

Wear protective safety glasses

In case of eye contact, flush eyes immediately with flowing water for 10–15 minutes, holding eyelids apart, and consult a doctor.

Read the instructions before use

Carefully follow this manual and the BactoSense PURE user manual.

Contents​

ItemQuantitySDS document numberUse
Alcohol Wipes10—Single use
Beads Solution1050106Single use
Reference Solution10—Single use
Sterile Solution10—Single use
Tube Holder1—Multiple uses

Discard single-use items according to local waste disposal regulations. Safety data sheets are available from bNovate Technologies SA upon request.

When to use the System Suitability Test Kit​

Perform an SST:

  • As part of quality assurance and validation, for example during operational qualification, performance qualification, cartridge change or servicing.
  • After storing the instrument for more than one month.
  • If there are indications that measurement results could be erroneous.

The procedure consists of three parts, performed in order: Optical Validation, Reference Validation and Sterile Validation. All test acceptance criteria must be met. The procedure takes approximately 2.5 hours.

Preparations​

  1. Verify that the SST Kit has not expired and has been stored under the specified conditions.
  2. Remove one alcohol wipe, one Beads Solution, one Reference Solution, one Sterile Solution and the Tube Holder.
  3. Make sure the manual sampling device is tightly installed on BactoSense PURE.
  4. Unpack the alcohol wipe. Wipe the sampling needle from top to bottom to remove contaminants. Let it dry for 30 seconds.
  5. Log in on BactoSense PURE with your personal username and password. Advanced or Admin permissions are required.
  6. Open Home → Maintenance → Instrument validation. Check Activate Instrument validation and press Save.
  7. Continue with Part A.

Part A: Optical Validation​

This tests the instrument's optical system by measuring a solution containing fluorescent particles.

  1. Attach the wash station to the manual sampling device.

  2. Open Home → Manual mode → Start. Select Clean Sampling Device with Washstation, with one replication. Press Next and start the protocol.

  3. After completion, shake the Beads Solution tube vigorously for 10 seconds, by hand or using a vortex mixer.

  4. Remove the caps from the Beads Solution tube and Tube Holder. Insert the opened tube into the holder. Remove the wash station and attach the holder containing the Beads Solution.

    Use the whole reagent tube

    Insert the entire reagent tube into the Tube Holder, rather than pouring its contents into the holder.

    Opening and inserting a single-use reagent tube into the Tube Holder

  5. Return to the protocol overview. Select Prime, with one replication. Press Next and start.

  6. After completion, return to the protocol overview. Select Beads Analysis, with one replication. Press Next and start.

  7. Review the results. All the following criteria must be met:

    TCC: 50,000–75,000/mL

    ChannelCV (%)Median
    FL1Less than 155.7–6.5
    FL2Less than 205.1–5.7
    SSCLess than 254.7–6.3

    Example optical validation results

  8. If the test passes, remove the holder, remove the Beads Solution tube, close its cap and discard it. Continue with Part B.

    If it fails, repeat, making sure the Beads Solution is well mixed. If failures persist, contact bNovate support.

Part B: Reference Validation​

This tests the instrument's automated staining process.

  1. Attach the wash station to the manual sampling device.

  2. Open Home → Manual mode → Start. Run Clean Sampling Device with Washstation, with one replication. Press Next and start.

  3. Remove the cap from a Reference Solution tube and insert the entire tube into the holder, as shown in Part A. After the cleaning protocol finishes, remove the wash station and attach the holder containing the Reference Solution.

  4. Return to the protocol overview. Run Prime, with one replication.

  5. After completion, return to the protocol overview. Select Water Analysis, with one replication. In Gates, select ICC Default Gate. Press Next and start.

  6. Compare the main bacteria clouds to the reference criteria:

    DotplotAcceptance criterion
    FL1 / FL2Main cloud above 4.5 on the FL1 axis and below 5 on the FL2 axis
    FL1 / SSCMain cloud above 2 and below 5.5 on the SSC axis

    Reference dotplots with the main bacteria clouds circled in orange

    Appearance of the clouds

    Depending on the cartridge and storage conditions, the clouds may differ from the example.

  7. If the test passes, remove the holder, remove the Reference Solution tube, close its cap and discard it. Continue with Part C.

    If it fails, repeat with a fresh Reference Solution. If failures persist, contact bNovate support.

Part C: Sterile Validation​

This verifies the cleanliness of the sampling device and fluidic system by measuring cell-free water.

  1. Attach the wash station to the manual sampling device.

  2. Open Home → Manual mode → Start. Run Clean Sampling Device with Washstation, with one replication. Press Next and start.

  3. Remove the cap from a Sterile Solution tube and insert the entire tube into the holder, as shown in Part A. After the cleaning protocol finishes, remove the wash station and attach the holder containing the Sterile Solution.

  4. Return to the protocol overview. Run Prime, with one replication.

  5. After completion, return to the protocol overview. Select Water Analysis, with one replication. Select ICC Default Gate, press Next and start.

  6. Review the result. The acceptance criterion is ICC less than 100/mL.

  7. If the test passes, remove the holder, remove the Sterile Solution tube, close its cap and discard it. Return the holder to the SST Kit, store the kit under the specified conditions, and attach the wash station to the manual sampling device.

    If it fails, perform biological decontamination using the Cleaning Kit. Repeat with a fresh Sterile Solution. If failures persist, contact bNovate support.

  8. Open Home → Maintenance → Instrument validation. Uncheck Activate Instrument validation and press Save.

You have successfully completed the System Suitability Test when all three parts have passed.